Get Adobe Flash player

HBX76

Ballast for Xenon and Mercury Short Arc lamps rated 50 to 100W


Key Benefits

  • Compact sizeCertification for medical applications
  • Built-in cooling fan
  • HBX76 igniter on board


Target Lamps

Manufacturer Type
OSRAM

HBO 50W/3
HBO 75W/3
HBO 100W/2
HBO 103W/2
HBO R 103/45
XBO 75W/2
XBO 100W OFR
XBO R 100W/ 45 (C) OFR
XBO R 101W/45 (C) OFR
XBO 103W/2

Ushio

SHL 50W DC
USH-103 D

Product Specification


Input Data

Nominal OperationNominalTolerance
Input Voltage AC (V) 100 - 240 90 - 264
Input Voltage DC (V)/ 100 - 300 90 - 340
System Wattage (W) 123 60 - 120*
Input Current (A) --- 0.6 - 1.4*
Line Frequency (Hz) 50/60 47 - 63
Power Factor (1)/ 1.0 0.93 - 1.0
Depends on lamp select. Presettable


Lamp Output Data

IgnitionNominalToleranceRemarks
Ignition Voltage (kV peak) +/-14 +/-12 - +/-16 Load capacity <20pF
Ignition Time (sec 1 0.9 - 1.1
Automatic restart counter (1) 5 --- attempts
Nominal Operation Nominal Tolerance
Lamp Voltage (W) 12.8 - 34 +/-5%
Lamp Wattage (W) 50, 75, 100* selctable 50, 75, 100
Lamp Current (A) --- I max = 7.5*
Depends on lamp selekt. Presetable


Lifetime Data

SymbolNominalToleranceRemarks
Ballast Lifetime t Life 25.000 >25.000 acc. to MIL HDBK for nominal operation


Miscellaneous

Nominal OperationNominalTolerance
Efficiency (1) 0.8 0.75 - 0.8
Ambient Temperature (°C) +25 +10 - +50
Max. Temperature at test poin(°C) +90 +85 - +95


Cooling Recommendations

The unit has 12V terminals for driving one or two fans. One is intended for the power supply and one for the lamp. By default, one in connected to the built - in power supply cooling fan. The combined maximum total output current for both outputs is 200 mA.This leaves 135mA for an external fan. Please note that this output voltage is only available when the lamp is in operation. Temperature of the power suppy should be exceed 70°C. Temperature overload protected by an internal temperature switch at 90°C.



Environmental Requirements

  • Storage Temperature Range -20°C to +50°C
  • Operation Temperature Range 0°C to +60°C
  • Humidity Range non condensing 20% to 95%
  • Altitude (operating) 0 ft to 10,000 ft


CAUTION:

SAFETY

Due to restrike capabilities of the power supply, the output voltage to the lamp can reach 15,000 volts. Ensure a minimum 15mm (<1/2") clearence between all lamp terminals to the power supply. All primary wiring must meet all local national safety regulation.

Specification subject to change without prior notice!