HBX300
Ballast for Xenon Short Arc lamps rated 150 to 300W
Key Benefits
- Compact size
- Certification for medical application
- Built-in cooling fan
Target Lamps
| Manufacturer | Type |
| OSRAM |
XBO 150 W/1 XBO 150 W/1 OFR XBO 150 W/4 XBO 150 W/CR OFR XBO 180 W/45 OFR XBO 181W/45C OFR XBO 250W OFR 18,0 A VS1 XBO 250W/4 18,0A VS1 XBO 450W/xxx +)
|
| Ushio |
SMH-200/D1 SMH-200/D2 SMH-270/D1 SMH-350/D2 +) SMH-600/D1 ++)
|
as cascaded operation together with HBX180
as cascaded operation consisting of two HBX300
Product Specification
Input Data
| NominalOperation | Nominal | Tolerance |
| Input Voltage AC (V) |
100 - 240 |
90 - 264 |
| Input Voltage DC (V) |
100 - 300 |
90 - 340 |
| System Wattage (W) |
--- |
150 - 350* |
| Input Current (A)/ |
--- |
0.6 - 1.4* |
| Line Frequency (Hz) |
50/60 |
47 - 63 |
| Power Faktor (1) |
1.0 |
0.93 - 1.0 |
*Depends on lamp select.
Lamp Output Data
| Ignition | Nominal | Tolerance | Remarks |
| Ignition Voltage with ZG 20 Xe |
+/-14 |
+/-12 - +/-16 |
Load capacity <20pF |
| Ignition Time (sec) |
1 |
0.9 - 1.1 |
|
| Automatic restart counter (1) |
5 |
--- |
attempts |
| Nominal Operation | Nominal | Tolerance |
| Lamp Voltage (V) |
10 - 29 |
+/-5%* |
| Lamp Wattage (W) |
150, 180, 200, 250, 300 |
+/-2% |
| Lamp Current (A) |
--- |
Depend on select |
*Depends on lamp select. Presetable
Lifetime Data
| | Symbol | Nominal | Tolerance | Remarks |
| Ballast Lifetime |
t Life |
25.000 |
>25.000 |
acc. to MIL HDBK for nominal operation |
Miscellaneous
| Nominal Operation | Nominal | Tolerance |
| Efficiency (1) |
0.83 |
0.8 - 0.9 |
| Ambient Temperature (°C) |
+25 |
+10 - +50 |
| Max. Temperature at test poin(°C) |
+80 |
+75 - +85 |
Cooling Recommendations
In all cases, the temperature at the temperature test point should be tested to ensure most reliable operations. Temperature should not exceed 80°C. Temperature overload protected by an internal temperature switch at 90°C.
Environmental Requirements>/h4>
- Storage Temperature Range -20°C to +50°C
- Operation Temperature Range 0°C to +60°C
- Humidity Range non condensing 20% to 95%
- Altitude (operating) 0 ft to 10,000 ft
CAUTION:
SAFETY:
Because of instant hot restrike, the output voltage on the lamp can reach values of up to +/- 15000 Volts (ZG 20 Xe). Please ensure minimum 15mm between lamp terminals to PE, to prevent arc to ground situation! Primary wiring has to meet national requirements for electric safety!
Specification subject to change without prior notice!