HBX300

Ballast for Xenon Short Arc lamps rated 150 to 300W



Key Benefits

  • Compact size
  • Certification for medical application
  • Built-in cooling fan


Target Lamps

Manufacturer Type
OSRAM

XBO 150 W/1
XBO 150 W/1 OFR
XBO 150 W/4
XBO 150 W/CR OFR
XBO 180 W/45 OFR
XBO 181W/45C OFR
XBO 250W OFR 18,0 A VS1
XBO 250W/4 18,0A VS1
XBO 450W/xxx +)

Ushio

SMH-200/D1
SMH-200/D2
SMH-270/D1
SMH-350/D2 +)
SMH-600/D1 ++)

as cascaded operation together with HBX180
as cascaded operation consisting of two HBX300


Product Specification



Input Data

NominalOperation Nominal Tolerance
Input Voltage AC (V) 100 - 240 90 - 264
Input Voltage DC (V) 100 - 300 90 - 340
System Wattage (W) --- 150 - 350*
Input Current (A)/ --- 0.6 - 1.4*
Line Frequency (Hz) 50/60 47 - 63
Power Faktor (1) 1.0 0.93 - 1.0
*Depends on lamp select.


Lamp Output Data

IgnitionNominalToleranceRemarks
Ignition Voltage with ZG 20 Xe +/-14 +/-12 - +/-16 Load capacity <20pF
Ignition Time (sec) 1 0.9 - 1.1
Automatic restart counter (1) 5 --- attempts
Nominal OperationNominalTolerance
Lamp Voltage (V) 10 - 29 +/-5%*
Lamp Wattage (W) 150, 180, 200, 250, 300 +/-2%
Lamp Current (A) --- Depend on select
*Depends on lamp select. Presetable


Lifetime Data

SymbolNominalToleranceRemarks
Ballast Lifetime t Life 25.000 >25.000 acc. to MIL HDBK for nominal operation


Miscellaneous

Nominal OperationNominalTolerance
Efficiency (1) 0.83 0.8 - 0.9
Ambient Temperature (°C) +25 +10 - +50
Max. Temperature at test poin(°C) +80 +75 - +85


Cooling Recommendations

In all cases, the temperature at the temperature test point should be tested to ensure most reliable operations. Temperature should not exceed 80°C. Temperature overload protected by an internal temperature switch at 90°C.



Environmental Requirements>/h4>
  • Storage Temperature Range -20°C to +50°C
  • Operation Temperature Range 0°C to +60°C
  • Humidity Range non condensing 20% to 95%
  • Altitude (operating) 0 ft to 10,000 ft


CAUTION:

SAFETY:

Because of instant hot restrike, the output voltage on the lamp can reach values of up to +/- 15000 Volts (ZG 20 Xe). Please ensure minimum 15mm between lamp terminals to PE, to prevent arc to ground situation! Primary wiring has to meet national requirements for electric safety!

Specification subject to change without prior notice!