HBX180

Ballast for Xenon Short Arc lamps rated 150 to 180W



Key Benefits

  • Compact size
  • Certification for medical application
  • Built-in cooling fan


Target Lamps

Manufacturer Type
OSRAM

XBO 150 W/1
XBO 150 W/1 OFR
XBO 150 W/4
XBO 150 W/CR OFR
XBO 180 W/45 OFR
XBO 181W/45C OFR

Ushio

SMH-200/D1
SMH-200/D2



Product Specification



Input Data

Nominal Operation Nominal Tolerance
Input Voltage AC (V) 100 - 240 90 - 264
Input Voltage DC (V) 100 - 300 90 - 340
System Wattage (W) 123 150 - 200*
Input Current (A) --- 0.6 - 1.4*
Line Frequency (Hz) 50/60 47 - 63
Power Factor (1) 1.0 0.93 - 1.0
Depends on lamp select.


Lamp Output Data

IgnitionNominalToleranceRemarks
Ignition Voltage with ZG 20 Xe +/-14 +/-12 - +/-16 Load capacity <20pF
Ignition Time (sec) 1 0.9 - 1.1 ---
Automatic restart counter (1) 5 --- attempts
Nominal OperationNominalTolerance
Lamp Voltage (V) 10 - 29 +/-5%*
Lamp Wattage (W) 150 - 180 Selectable by mode Sw.
Lamp Current (A) --- Depend on select
*Depends on lamp select. Presetable


Lifetime Data

SymbolNominalToleranceRemarks
Ballast Lifetime t Life 25.000 >25.000 acc. to MIL HDBK for nominal operation


Miscellaneous

Nominal OperationNominalTolerance
Efficiency (1) 0.83 0.8 - 0.9
Ambient Temperature (°C) +25 +10 - +50
Max. Temperature at test poin(°C) +80 +75 - +85/td>


Cooling Recommendations

In all cases, the temperature at the temperature test point should be tested to ensure most reliable operations. Temperature should not exceed 80°C. Temperature overload protected by an internal temperature switch at 90°C.


Environmental Requirements

  • Storage Temperature Range -20°C to +50°C
  • Operation Temperature Range non condensing -20°C to +60°C 20% to 95%
  • Humidity Range 0 ft to 10,000 ft
  • Altitude (operating) 0 ft to 40,000 ft


CAUTION:



SAFETY:

Because of instant hot restrike, the output voltage on the lamp can reach values of up to +/- 15000 Volts (ZG 20 Xe) or 30.000 Volts at the unsymmetrical ZG20Xe-S! Please ensure minimum 15mm between lamp terminals to PE, to prevent arc to ground situation! Primary wiring has to meet national requirements for electric safety!



Specification subject to change without prior notice!