HBX180
Ballast for Xenon Short Arc lamps rated 150 to 180W
Key Benefits
- Compact size
- Certification for medical application
- Built-in cooling fan
Target Lamps
| Manufacturer | Type |
| OSRAM |
XBO 150 W/1 XBO 150 W/1 OFR XBO 150 W/4 XBO 150 W/CR OFR XBO 180 W/45 OFR XBO 181W/45C OFR
|
| Ushio |
SMH-200/D1 SMH-200/D2
|
Product Specification
Input Data
| Nominal Operation | Nominal | Tolerance |
| Input Voltage AC (V) |
100 - 240 |
90 - 264 |
| Input Voltage DC (V) |
100 - 300 |
90 - 340 |
| System Wattage (W) |
123 |
150 - 200* |
| Input Current (A) |
--- |
0.6 - 1.4* |
| Line Frequency (Hz) |
50/60 |
47 - 63 |
| Power Factor (1) |
1.0 |
0.93 - 1.0 |
Depends on lamp select.
Lamp Output Data
| Ignition | Nominal | Tolerance | Remarks |
| Ignition Voltage with ZG 20 Xe |
+/-14 |
+/-12 - +/-16 |
Load capacity <20pF |
| Ignition Time (sec) |
1 |
0.9 - 1.1 |
--- |
| Automatic restart counter (1) |
5 |
--- |
attempts |
| Nominal Operation | Nominal | Tolerance |
| Lamp Voltage (V) |
10 - 29 |
+/-5%* |
| Lamp Wattage (W) |
150 - 180 |
Selectable by mode Sw. |
| Lamp Current (A) |
--- |
Depend on select |
*Depends on lamp select. Presetable
Lifetime Data
| | Symbol | Nominal | Tolerance | Remarks |
| Ballast Lifetime |
t Life |
25.000 |
>25.000 |
acc. to MIL HDBK for nominal operation |
Miscellaneous
| Nominal Operation | Nominal | Tolerance |
| Efficiency (1) |
0.83 |
0.8 - 0.9 |
| Ambient Temperature (°C) |
+25 |
+10 - +50 |
| Max. Temperature at test poin(°C) |
+80 |
+75 - +85/td> |
Cooling Recommendations
In all cases, the temperature at the temperature test point should be tested to ensure most reliable operations. Temperature should not exceed 80°C. Temperature overload protected by an internal temperature switch at 90°C.
Environmental Requirements
- Storage Temperature Range -20°C to +50°C
- Operation Temperature Range non condensing -20°C to +60°C 20% to 95%
- Humidity Range 0 ft to 10,000 ft
- Altitude (operating) 0 ft to 40,000 ft
CAUTION:
SAFETY:
Because of instant hot restrike, the output voltage on the lamp can reach values of up to +/- 15000 Volts (ZG 20 Xe) or 30.000 Volts at the unsymmetrical ZG20Xe-S! Please ensure minimum 15mm between lamp terminals to PE, to prevent arc to ground situation! Primary wiring has to meet national requirements for electric safety!
Specification subject to change without prior notice!